Sign in →

Test Code GLUF Fasting Glucose, LAB81

Important Note

Order to be used for fasting outpatient services.

Performing Laboratory

St. Luke's Hospital Cedar Rapids

Specimen Requirements

Container/Tube:

Preferred: Green-top (lithium heparin) tube

Acceptable:Serum, K2‑EDTA, NaF/Na2EDTA, KF/Na2EDTA, NaF/K‑Oxalate and NaF/citrate/Na2‑EDTA are acceptable. Plasma or serum samples without preservative (NaF) should be separated from the cells or clot within half an hour of being drawn.

Specimen Volume: Minimum volume of 1.5 µL plus a dead volume

Stability: Separated specimens are stable for as long as 8 hours at 25°C and up to 72 hours at 2-8⁰C.

Reference Values

0-2 days:  55-65 mg/dL

>2 days:  70-99 mg/dL


Critical Values

Serum/Plasma

Outpatient

Critical low

0 – 2 days

45 mg/dL

Always call

 

3 days – 12 years

59 mg/dL

Always call

 

>12 years

59 mg/dL

Call during business hours

Critical high

0 – 2 days

200 mg/dL

Always call

 

>2 days

>400 mg/dL

Always call

Day(s) Test Set Up

Monday through Sunday

Useful For

Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Methodology

Hexokinase/Ultraviolet

Test Classification and CPT Coding

82947

Report Available

STAT within 1 hour of receipt

Routine within 2 hours of receipt