Test Code PROTC Protein C Activity, Plasma

Additional Codes

 Test ID

PROTC

EPIC Test ID

LAB489

Specimen Requirements

Container/Tube Type: 

Light Blue (Citrate)

Collection Information: 

Promptly after collection, spin down and place 1 mL of platelet-poor plasma in plastic vial. (Glass vial is not acceptable.) Freeze specimen immediately at ≤ -40° C, if possible.

Platelet Poor Plasma
Centrifuge Specimen

Recommended centrifuge speed is 1500 x G for at least 15 minutes. Be careful that the tubes are evenly balanced when placing them into the centrifuge. When the centrifuge stops, carefully remove and rack tubes. Be aware that an unbalanced centrifugation, speed too slow, or centrifugation for too short of a spin time may all lead to an increase number of cells in the plasma which could contribute to a clotted specimen or erroneous results.
Remove Plasma
To remove plasma, start from the top, gently draw specimen into pipette as you go further down tube. Leaving approximately 0.5mL of plasma will insure that you do not disturb the buffy coat and cell layer.
Second Spin
Plasma separated from cells goes back into the centrifuge for a second spin. Balance and centrifuge the tubes. Failure to perform a second spin will result in plasma containing too many platelets which interfere with some coagulation testing and is a contributing factor for a clotted specimen.
Transfer Plasma to Aliquot Tubes
Pipette plasma from second spin tube into labeled aliquot tubes without disturbing the bottom 0.5mL.
Specimen Storage
Promptly freeze specimens after processing.

Minimum Sample Volume

0.5 mL

Specimen Stability

Refrigerated Stability: 

NOT Acceptable

Frozen Stability: 

14 days

Ambient Stability: 

NOT Acceptable

Report Availibility

Performed: 

Once per day Tuesday and Thursday, Dayshift Only (Specimen must be in lab prior to 10 am)

Methodology

Chromogenic Assay

Unacceptable Conditions

Hemolyzed or lipemic specimens.